ISO 14644 Sampling Location Calculator

Work out the minimum number of sampling locations for cleanroom classification from the room area, using Table A.1 of ISO 14644-1:2015. Enter the area in m² or ft²; the ISO class does not change the number of locations.

Cleanroom classification · ISO 14644-1

ISO 14644 Sampling Location Calculator

Runs in your browser
Cleanroom

Normally the floor (horizontal plan) area of the cleanroom or clean zone. For unidirectional airflow, ISO 14644-1:2015 allows the cross-section of the moving air, perpendicular to the airflow. The ISO class does not change the number of locations.

Minimum sampling locations

NL = –

Enter the cleanroom area.

This gives the minimum for classification only. It is not an approved sampling plan and does not replace your risk assessment, ISO 14644-2 monitoring or GMP environmental monitoring requirements.

What is ISO 14644-1 sampling location determination?

To classify a cleanroom, airborne particle concentrations are measured at a number of points and each point must meet the limit for the target ISO class. ISO 14644-1 sets the minimum number of these sampling locations so that the result represents the whole room. In the 2015 edition, the number in Table A.1 is chosen so that, if every location complies, there is about 95% confidence that at least 90% of the cleanroom area meets the class limit.

For the ISO classes themselves (limits for ISO 1 to ISO 9, the class formula, occupancy states and retest intervals), see the guide to ISO 14644-1 cleanroom classification.

How many sampling locations are required?

Under ISO 14644-1:2015, the minimum number of locations (NL) depends only on the area. It does not depend on the ISO class. For rooms up to 1,000 m², NL is read from Table A.1. If the area falls between two values, the larger value is used.

Area (m²) up toNLArea (m²) up toNLArea (m²) up toNL
21521014819
42561115620
63641219221
84681323222
105721427623
246761535224
2871041643625
3281081763626
369116181,00027
Minimum sampling locations versus cleanroom area, ISO 14644-1:2015 Table A.1. Above 1,000 m², use Formula (A.1).

For rooms larger than 1,000 m², Formula (A.1) applies: NL = 27 × (A ÷ 1,000), rounded up to a whole number. The area is normally the horizontal plan area. In unidirectional airflow, it may be taken as the cross-section of the moving air perpendicular to the airflow.

How to calculate minimum sampling locations: worked example

A cleanroom measures 4.0 m × 3.0 m = 12 m².

  1. Find NL: in Table A.1, 12 m² falls between the 10 m² and 24 m² rows, so the larger row (≤ 24 m²) is used: NL = 6.
  2. Divide the room into 6 sections of equal area: 12 m² ÷ 6 = 2 m² per section. Here the room is split into 3 columns × 2 rows, so each section is 1.33 m × 1.5 m.
  3. Place one location in each section at a point that represents that section, usually at work height and near the centre unless equipment, airflow or the activity justify another point. This gives locations L1 to L6.
12 m² cleanroom divided into 6 equal sectionsA 4.0 m by 3.0 m cleanroom split into 3 columns and 2 rows, giving six sections of 2 m² each, with one sampling location L1 to L6 near the centre of each section.Section 1 · 2 m²L1Section 2 · 2 m²L2Section 3 · 2 m²L3Section 4 · 2 m²L4Section 5 · 2 m²L5Section 6 · 2 m²L64.0 m3.0 m1.33 m1.33 m1.33 m1.5 m1.5 mSampling location (one per section)Section boundary
Example: a 12 m² cleanroom needs 6 sampling locations (ISO 14644-1:2015 Table A.1). The room is divided into 6 sections of 2 m² each, with one location per section.

Other layouts are acceptable, for example 6 strips of 0.67 m × 3.0 m, as long as the sections have equal area and each location represents its section. For a larger room of 2,500 m², NL = 27 × 2,500 ÷ 1,000 = 67.5, rounded up to 68 locations.

Minimum sample volume and sampling duration

  • Number of locations: set by the area (Table A.1).
  • Minimum single sample volume (Vs): Vs = (20 ÷ Cn,m) × 1,000 litres, where Cn,m is the class limit (particles/m³) for the largest considered particle size. It is the volume in which 20 particles would be counted if the concentration were exactly at the class limit. The volume at each location must be at least 2 L, and each location must be sampled for at least 1 minute.
  • Flow rate: a property of the particle counter, for example 28.3 L/min (1 ft³/min), 50 L/min or 100 L/min.
  • Sampling duration: required volume ÷ flow rate, but never less than 1 minute.

The volume is set by the class limit at the largest particle size considered. For ISO Class 5 (the cleanest class used in pharmaceutical manufacturing), ≥ 0.5 µm (3,520 particles/m³) needs 5.68 L and ≥ 1 µm (832 particles/m³) needs about 24.0 L per location. When the EU GMP Grade A limit of 29 particles/m³ at ≥ 5.0 µm is included, about 690 L is needed. ISO 14644-1:2015 allows a sequential sampling procedure (Annex D) when large volumes are needed. EU GMP Annex 1 grades are not the same as ISO classes; see cleanroom grades A, B, C and D vs ISO classes.

Worked example: EU GMP Grade A zone

A Grade A filling zone measures 2.5 m × 1.6 m = 4 m². Its limits are 3,520 particles/m³ at ≥ 0.5 µm and 29 particles/m³ at ≥ 5.0 µm (EU GMP Annex 1).

StepCalculationResult
Sampling locationsTable A.1: 4 m² is within the row “≤ 4 m²”2 locations
Vs at ≥ 0.5 µm(20 ÷ 3,520) × 1,0005.68 L
Vs at ≥ 5.0 µm(20 ÷ 29) × 1,000689.7 L
Volume per locationLargest considered size (5.0 µm) decidesat least 690 L
Time per location if the particle counter flow rate is 28.3 L/min689.7 ÷ 28.324 min 23 s
Time per location if the particle counter flow rate is 50 L/min689.7 ÷ 5013 min 48 s
Time per location if the particle counter flow rate is 100 L/min689.7 ÷ 1006 min 54 s

The 5.0 µm limit drives the volume: about 121 times more air than the 0.5 µm limit alone would need. ISO 14644-1:2015 Table 1 gives no ISO Class 5 limit at 5.0 µm, so this limit comes from the GMP requirement, not from ISO 14644-1. Check Annex 1 and your procedure for any further minimum sample volume.

How to select sampling locations in a cleanroom

ISO 14644-1:2015 asks for the cleanroom to be divided into NL sections of equal area, with one location in each section at a point representative of that section. Beyond that, these practical points help:

  • Sample at work height, or the height relevant to the activity, unless a risk assessment justifies another height.
  • Consider airflow patterns, equipment, critical operations, personnel movement and likely contamination sources.
  • Add extra locations at critical points; they are on top of the minimum and are recorded in the report.
  • Avoid clustering locations together, and in non-unidirectional airflow avoid points directly under an undiffused supply filter.
  • For irregular rooms, draw the sections on a floor plan and document the reasoning.

Common errors during cleanroom sampling

  • Wrong area unit: Table A.1 is in m². 1 ft² = 0.0929 m², so 1,000 ft² is about 92.9 m² (16 locations), not 1,000 m².
  • Confusing volume with time: the volume comes from the class limit; the time comes from volume ÷ flow rate, with a 1-minute minimum.
  • Wrong flow rate: using a nominal rate that does not match the counter’s calibrated flow changes the time and the volume sampled.
  • Poor distribution: placing all locations near the door or in one corner does not represent the room.
  • Treating the minimum as a plan: the calculator gives a minimum count, not an approved, risk-based sampling plan.

Frequently Asked Questions (FAQs)

How many sampling locations are required for an ISO Class 7 cleanroom?

It depends on the area, not the class. An ISO Class 7 room of 30 m² needs 8 locations, and one of 100 m² needs 16, under ISO 14644-1:2015 Table A.1.

What sample volume is needed for an EU GMP Grade A zone?

If ≥ 5.0 µm is the largest particle size considered, with a limit of 29 particles/m³, Vs = (20 ÷ 29) × 1,000 = about 690 L per location. That is about 7 minutes with a 100 L/min counter. If only ≥ 0.5 µm (3,520 particles/m³) is considered, 5.68 L is enough, so the 1-minute minimum applies.

Is the number of sampling locations based only on cleanroom area?

The minimum is based only on area. Additional locations may be added for critical points, unusual layouts or risk, and they are recorded in the report.

What is the minimum sample volume for cleanroom classification?

Vs = (20 ÷ Cn,m) × 1,000 litres for the largest considered particle size, with at least 2 L and at least 1 minute per location. There is no single volume that fits every class.

How does particle counter flow rate affect sampling duration?

Duration = required volume ÷ flow rate, with a 1-minute minimum. For 689.7 L (EU GMP Grade A, 29 particles/m³ at ≥ 5.0 µm), a 28.3 L/min counter needs about 24 min 23 s, and a 100 L/min counter needs about 6 min 54 s. For 24.0 L (ISO Class 5 at ≥ 1 µm), both counters need the 1-minute minimum.

Can this calculator be used for pharmaceutical cleanroom qualification?

It gives the ISO 14644-1 minimum for classification. Pharmaceutical qualification also follows GMP requirements, such as EU GMP Annex 1, and a documented risk assessment. These may require more locations or different sample volumes.

References

  • ISO 14644-1:2015, Cleanrooms and associated controlled environments, Part 1: Classification of air cleanliness by particle concentration (Annex A, Table A.1, Formula A.1, and Table 1).
  • ISO 14644-2:2015, Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration.
  • EU GMP Annex 1 (2022), Manufacture of Sterile Medicinal Products.

This calculator is an educational tool. It is not endorsed by ISO or any regulator. The ISO standards are copyrighted and not freely available; the table values used here were checked against published reproductions of the 2015 tables. Always verify against your licensed copy of the standard, your GMP requirements and your approved procedures.