This calculator tells you how many containers to submit for a USP <71> sterility test and how much product is tested from each one. Choose the product type, enter the batch size and the fill of one container, and it applies the chapter’s two sampling tables: the minimum number of items for the batch size, and the minimum quantity for each culture medium.
How the sample size is worked out
USP <71> uses two tables. One sets the minimum number of items to test, based on the type of product and the number of containers in the batch. The other sets the minimum quantity of product to put into each culture medium, based on the fill of one container. The test uses two media, Fluid Thioglycollate Medium and Soybean–Casein Digest Medium.
- If one container holds enough product for both media, the number of items from the first table covers both media together.
- If one container does not hold enough for both media (for example, a 1 mL syringe, where each medium needs 1 mL), twice that number of containers is needed.
- Percentages are rounded up to the next whole container.
Minimum number of items to test
| Product | Batch size | Minimum number of items |
|---|---|---|
| Parenteral preparations | Not more than 100 containers | 10% or 4 containers, whichever is greater |
| More than 100, not more than 500 | 10 containers | |
| More than 500 | 2% or 20 containers, whichever is less | |
| Large-volume parenterals | Any | 2% or 10 containers, whichever is less |
| Antibiotic solids, pharmacy bulk packages | Under 5 g per container | 20 containers |
| 5 g or more per container | 6 containers | |
| Ophthalmic and other non-injectable preparations | Not more than 200 containers | 5% or 2 containers, whichever is greater |
| More than 200 containers | 10 containers | |
| Single-dose containers | Same as parenteral preparations | |
| Medical devices and surgical materials | Same batch-size bands as parenterals | 10% or 4; 10; or 2% or 20 articles |
| Catgut and sutures for veterinary use | Any | 2% or 5 packages, whichever is greater, up to 20 |
| Bulk solid products | Up to 4 containers | Each container |
| More than 4, not more than 50 | 20% or 4 containers, whichever is greater | |
| More than 50 | 2% or 10 containers, whichever is greater |
Minimum quantity for each medium
| Contents of one container | Minimum quantity for each medium |
|---|---|
| Liquid, less than 1 mL | Whole contents |
| Liquid, 1 to 40 mL | Half the contents, not less than 1 mL |
| Liquid, more than 40 mL up to 100 mL | 20 mL |
| Liquid, more than 100 mL | 10% of the contents, not less than 20 mL |
| Antibiotic liquids | 1 mL |
| Soluble preparations, insoluble preparations, creams and ointments | Enough to give not less than 200 mg |
| Solid, less than 50 mg | Whole contents |
| Solid, 50 mg to less than 300 mg | Half the contents, not less than 50 mg |
| Solid, 300 mg to 5 g | 150 mg |
| Solid, more than 5 g | 500 mg |
| Catgut and sutures for veterinary use | 3 sections of strand, each 30 cm long |
| Surgical dressings, cotton and gauze | 100 mg per package |
| Individually packaged single-use materials | The whole device |
| Other medical devices | The whole device, cut into pieces or taken apart |
Worked examples
400 vials of 50 mL injection. A batch of more than 100 but not more than 500 containers needs 10 containers. A 50 mL fill needs 20 mL for each medium, so one vial (40 mL for both media) is enough. Submit 10 vials; 400 mL is tested in total.
600 prefilled syringes of 1 mL. More than 500 containers needs 2% or 20, whichever is less: 2% of 600 is 12. A 1 mL fill needs 1 mL for each medium, so one syringe cannot supply both media and the number is doubled. Submit 24 syringes.
150 bottles of a 5 mL multi-dose eye drop. Not more than 200 containers needs 5% or 2, whichever is greater: 5% of 150 rounds up to 8. Each medium needs half of 5 mL (2.5 mL), so one bottle covers both media. Submit 8 bottles.
Before you send samples to the laboratory
- The numbers above are the minimum for the test itself. Method suitability testing, and a possible retest, need additional units. Agree the total with the laboratory in advance.
- The test method (membrane filtration or direct inoculation) is chosen and validated for each product; it does not change the minimum number of items.
- Samples should represent the whole batch, for example from the beginning, middle and end of filling, as set out in your sampling procedure.
For the test method itself, media, incubation and how results are read, see sterility testing (USP <71>) and growth promotion testing of culture media.
Frequently Asked Questions (FAQs)
How many vials are needed for a USP 71 sterility test?
It depends on the batch size and the fill. For parenteral batches: 10% or 4 containers (whichever is greater) up to 100 containers, 10 containers for 101 to 500, and 2% or 20 (whichever is less) above 500. The number doubles when one container cannot supply both media.
Why is the number of containers sometimes doubled?
Each container must supply the minimum quantity for both culture media. When one container holds less than twice the minimum quantity, for example a 1 mL ampoule, twice the number of containers is tested.
How much is used from each container?
For liquids between 1 and 40 mL, half the contents (at least 1 mL) goes into each medium. For 40 to 100 mL, 20 mL per medium. Above 100 mL, 10% of the contents, at least 20 mL. Small fills under 1 mL use the whole contents.
Is the sample size the same in the European and Japanese Pharmacopoeias?
The sterility test chapter is harmonised between USP <71>, Ph. Eur. 2.6.1 and JP 4.06, so the sampling tables are the same. Some text in each pharmacopoeia is not harmonised, so check the version that applies to your product.
Does a passing sterility test prove the whole batch is sterile?
No. The test examines a small sample and can detect only gross contamination. Sterility assurance comes from validated sterilisation or aseptic processing and environmental control; the sterility test is one part of batch release.
References
- United States Pharmacopeia. General Chapter <71> Sterility Tests. Tables 2 and 3.
- European Pharmacopoeia. Chapter 2.6.1 Sterility.
- Japanese Pharmacopoeia. General Test 4.06 Sterility Test.
- ICH Q4B Annex 8. Evaluation and Recommendation of Pharmacopoeial Texts for Use in the ICH Regions on Sterility Test General Chapter.
This calculator is a working aid. For batch release, follow the current official chapter, your product specification and approved procedures.