Work out the minimum air volume and sampling time for each location when classifying a cleanroom to ISO 14644-1:2015. Pick an EU/PIC/S GMP grade or enter any ISO class, tick the particle sizes, and enter your particle counter’s flow rate.
How many locations? That depends on the room area. Use the ISO 14644 sampling location calculator.
Method: ISO 14644-1:2015, Annex A. Vs = (20 / Cn,m) × 1000 L; each sample at least 2 L and at least 1 minute. Class limits from Table 1; intermediate classes from Cn = 10N × (0.1/D)2.08, rounded to 3 significant figures. Use for planning; your SOP and the standard text govern.
How the minimum sample volume is calculated
ISO 14644-1 Annex A sets the single-sample volume so that at least 20 particles would be counted if the air were exactly at the class limit for the largest particle size considered:
Vs = (20 / Cn,m) × 1,000 litres
Each sample must also be at least 2 litres and run for at least 1 minute. The required volume is therefore the largest of the three values.
| Case | Class limit (Cn,m) | Minimum volume | Run time at 28.3 L/min (1 CFM) | Run time at 100 L/min |
|---|---|---|---|---|
| Grade A / ISO 5, ≥0.5 µm | 3,520 /m³ | 5.68 L → 1-minute rule governs | 1 min (28.3 L) | 1 min (100 L) |
| ISO 5, ≥5.0 µm included | 29 /m³ | 689.66 L | 24.37 min → 25 min (707.5 L) | 6.90 min → 7 min (700 L) |
| ISO 4, ≥1.0 µm | 83 /m³ | 240.96 L | 8.51 min → 9 min (254.7 L) | 2.41 min → 3 min (300 L) |
| ISO 6, ≥5.0 µm | 293 /m³ | 68.26 L | 2.41 min → 3 min (84.9 L) | 0.68 min → 1 min (100 L) |
| Grade C at rest / ISO 7, ≥5.0 µm | 2,930 /m³ | 6.83 L → 1-minute rule governs | 1 min (28.3 L) | 1 min (100 L) |
| Grade D at rest / ISO 8, ≥5.0 µm | 29,300 /m³ | 0.68 L → 2 L and 1-minute rules govern | 1 min (28.3 L) | 1 min (100 L) |
Worked example (Grade A with ≥5.0 µm, 100 L/min counter): Vs = (20 / 29) × 1,000 = 689.66 L. Time = 689.66 ÷ 100 = 6.90 min. Counters are set in whole minutes, so round up to 7 minutes, which collects 700 L and meets the minimum. Rounding down to 6 minutes (600 L) would fail the requirement. With a 1 CFM counter the same sample needs 689.66 ÷ 28.3 = 24.37 min, so set 25 minutes.
Points to watch
- Use the class limit for the largest size you are classifying at; small sizes rarely govern.
- Including ≥5.0 µm at ISO 5 makes each sample about 24 minutes at 28.3 L/min. EU/PIC/S Annex 1 does not require ≥5.0 µm for Grade A classification unless the contamination control strategy calls for it.
- Higher flow counters (50 or 100 L/min) shorten long samples but the 1-minute minimum still applies.
- This is for classification. Routine monitoring volumes are set in your monitoring plan (ISO 14644-2 and Annex 1).
Next step: find how many locations you need with the ISO 14644 sampling location calculator, then see the limits and monitoring rules in ISO 14644-1 cleanroom classification.
Reference
ISO 14644-1:2015, Cleanrooms and associated controlled environments, Part 1: Classification of air cleanliness by particle concentration, Annex A. European Commission, EU GMP Annex 1 (2022), Table 1.
