Lysate Sensitivity Confirmation Calculator (Gel-Clot BET)

Record the tube readings from the test for confirmation of labeled lysate sensitivity in the gel-clot bacterial endotoxins test, and the calculator works out each replicate endpoint, the geometric mean endpoint concentration (antilog Σe/f) and whether the labeled sensitivity λ is confirmed (0.5λ to 2λ), as described in USP General Chapter <85>. It also checks the validity condition: the lowest concentration (0.25λ) must be negative in all four replicates.

Gel-clot BET · USP <85> preparatory test

Lysate Sensitivity Confirmation Calculator

Runs in your browser
Lysate

The sensitivity stated on the lysate label, in EU/mL.

Tube results

Tap each tube to record the reading after inversion: +Ve firm gel (clot) · −Ve no gel. E. conc. is the endpoint of each set.

Geometric mean endpoint concentration

–

Enter λ and read every tube.

Formula

GM = antilog Σef

e = log10 of each replicate endpoint (EU/mL) · f = number of replicates (4). Accept when 0.5λ ≤ GM ≤ 2λ.

USP <85>: confirm λ in four replicates when a new lysate batch is used or test conditions change. The test is valid when the lowest concentration (0.25λ) is negative in all replicates. The endpoint is the smallest concentration that clots the lysate.

Learn how lysate sensitivity is confirmed

Want to check the reasoning behind the result? Our guide Lysate Sensitivity Confirmation for Gel-Clot BET: A Practical Guide explains which concentrations to prepare, how to read a tube, how the endpoint and geometric mean are found, validity checks and common mistakes.

This calculator is an educational tool. It is not approved or endorsed by USP or any regulator. Always verify results against the current pharmacopoeial text, the reagent manufacturer’s instructions and your validated laboratory procedures.